DO NOT immerse the Oximeter in liquid. If it needs cleaning, wipe the surface with medical
alcohol using a soft material. Do not spray any liquid directly onto the device.
When cleaning the device with water, ensure that the temperature of the water is lower than
60℃ .
Thin or cold fingers will affect the normal measure of the patient's SpO
sure to clip a thick finger such as the thumb or middle finger deeply enough into the probe.
The update period of data is less than 5 seconds. This is changeable according to different
individual pulse rates.
The waveform is normalized. Please read the measured value when the waveform on-screen
is uniform and steady-going. This measured value is the optimal value and this waveform is
the standard one.
If abnormal conditions appear on the screen during the test process, pull out the finger and
reinsert it to restore normal use.
The device has a lifespan of three years after the first electrified use.
The wrist strap attached to the device is made from non-allergenic material. If you are
sensitive to the wrist strap, stop using it. In addition, pay attention to the use of the wrist
strap. To avoid harm, DO NOT wear it around the neck.
This device has an alarm function.
Provided the alarm function is turned on, an alarm will automatically go off if the measured
data goes beyond the highest or lowest limit.
The alarm function can either be paused or turned off (default setting). The function can be
turned back on through menu operation. Please refer to Chapter 6.1 for more information on
the alarm function.
The device may not work for all patients. If you are unable to achieve stable readings,
discontinue use.
A flexible circuit connects the two parts of the device. DO NOT twist or pull on the
connection.
2. Overview
The peripheral capillary oxygen saturation (SpO
total hemoglobin (Hb) in the blood, or the oxygen concentration in the blood. It is an important bio-
parameter for respiration. A number of diseases relating to the respiratory system may cause a decrease
of SpO
in the blood. Furthermore, some other causes such as the malfunction of the body's self-
2
adjustment, damage during surgery, and injuries caused by some medical examinations could also lead to
a low oxygen supply in the human body. As a consequence, symptoms such as vertigo, impotence and
vomiting would appear. Serious symptoms might endanger the patient's life. Therefore, prompt
information of patients' SpO
importance in the clinical medical field.
The Pulse Oximeter features a small size, low power consumption, convenient operation and portability.
The patient only needs to put one of their fingers into a probe for diagnosis and a display screen will
directly show the measured value of pulse oxygen saturation with high accuracy.
2.1. Features
A. Operation of the product is simple and convenient.
B. The product is small in size and light in weight, making it convenient to carry.
C. The device has low power consumption.
TORONTEK-E400 Pulse Oximeter
) is the percentage of oxyhemoglobin (HbO
2
is of great help for doctors to discover potential dangers and is of great
2
2
and pulse rate. Be
2
) in the
2
Need help?