pressure cuff, or receiving intravenous injection.
D. Do not fix the SpO
inaccurate measure of the SpO
E. Excessive ambient light may affect the measuring result. This includes fluorescent lamps, dual
ruby lights, infrared heaters, direct sunlight etc.
F. Strenuous action of the patient or extreme electrosurgical interference may also affect accuracy.
G. Patient cannot use nail enamel or other makeup.
H. Please clean and disinfect the device after operating it according to the User Manual (7.1).

6.3. Clinical Restrictions

A. As the measurement is taken on the basis of arteriole pulse, substantial pulsating blood flow of
the subject is required. For a subject with weak pulse due to shock, low ambient/body
temperature, major bleeding, or use of vascular contracting drugs, the SpO
will decrease. In this case, the measurement will be more sensitive to interference.
B. For those with a substantial amount of staining dilution drugs (such as methylene blue, indigo
green and acid indigo blue), carboxyhemoglobin (COHb), methionine (Me+Hb) or thiosalicylic
hemoglobin, and some with an icterus problem, the SpO
inaccurate.
C. Drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor
blamed for serious errors of SpO
D. As the SpO
value serves as a reference value for judgment of anemic anoxia and toxic anoxia,
2
some patients with serious anemia may also report good SpO

7. Maintenance, Transportation and Storage.

7.1. Cleaning and Disinfection

Use alcohol wipes to disinfect the device and leave to air-dry. Otherwise, clean the device with a
soft cloth.

7.2. Maintenance

A. Please clean and disinfect the device before use according to the User Manual (7.1).
B. Please recharge the battery when the screen shows the low battery icon:
C. Recharge the battery soon after discharge. The device should be recharged every six months if it
is not regularly used. This will extend the battery life.
D. Users are advised to calibrate the device termly (or according to the calibrating program of the
hospital). It can also be performed at the state-appointed agent or you can contact us for
calibration.

7.3. Transportation and Storage

A. The packed device can be transported by ordinary conveyance or according to transport
contract. DO NOT transport the device with toxic or corrosive materials.
B. The best storage environment of the device is -40ºC~60ºC ambient temperature and not higher
than 95% relative humidity.
TORONTEK-E400 Pulse Oximeter
sensor with adhesive or else it may result in venous pulsation and
2
and pulse rate.
2
measurements.
2
10
waveform (PLETH)
2
determination by this monitor may be
2
measurement.
2
.