4.6 Deposit Options
Deposit Options is used to define the deposition potential for stripping experiments.
The default for this parameter is the initial potential for the potential scan (Initial E).
If some other potential is required, then the Deposit E option should be selected, and
the required potential should be entered in the Deposit E field (Figure 4-5).
Figure 4-5. Deposit Options dialog box.
4.7 Stabilizing Capacitor
A stabilizing capacitor is only required when using iR compensation, or for Bulk
Electrolysis (BE) experiments. In both these cases, the Auto option switches a 0.1 µF
or 1 µF capacitor into the circuit. The Manual option generally only needs to be used
when using the Manual iR option (Figure 4-6). The Manual option should be used
with caution, since the addition of capacitors can affect the frequency response of the
system.
Figure 4-6. Stabilizing Capacitor dialog box.
4-4