Hold-Ramp-Step (Hr) - BAS 100B/W Instruction Manual

Electrochemical workstation
Table of Contents

Advertisement

10.9.3 Hold-Ramp-Step (HR)

HR allows the programming of up to 12 time and/or potential based segments. It is
used to build custom waveforms for special applications.
Odd-numbered segments are chronoamperometric (current vs. time at a fixed
potential) and even-numbered segments are voltammetric (current vs. applied
potential ramp).
Mode = HR
General Parameters
Time Domain = (sec, µV/sec) or (msec, mV/sec)
Data Set = 1, 3, 5, 7, 9, 11
Hold E (mV) = 0
Hold Time = 0
Data Sets 2, 4, 6, 8, 10, 12
Initial E (mV) = 0
Final E (mV) = 0
Scan Rate = 100
Sensitivity = 1 µA/V
Specific Parameters
Quiet Time (sec) = 2
Graphics Menu
Single Graph displays the current vs. time plot for the Data Set selected
using Data Sets in Graphics Options.
Analysis Menu
The Auto option for Results Graph displays the current vs. potential curve
for the even-numbered Data Set selected using Data Sets in Graphics or
Results Options.
-3276 to 3276
0 to 65535
-3276 to 3276
-3276 to 3276
1 to 51200
100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
0 to 65535
10-66