CO2 Sample Draw Device Manual
Available in 1%, 30%, and 100% CO2 concentration ranges with optional data logging 

Introduction

Thank you for your purchase of the AQ500 CO2 Sampler. Please read this manual carefully before using your new sampler to ensure 
correct operating procedures are followed. The procedures outlined in this manual are vital to the accuracy, longevity, and correct 
usage of this product.  

Warnings

PAY SPECIAL ATTENTION TO BATTERY POLARITY WHEN INSERTING INTO UNIT. INSERTING THE BATTERIES BACKWARDS MAY CAUSE 
PERMANENT DAMAGE TO THE UNIT. ONLY USE THE PROVIDED POWER SUPPLY WITH THE UNIT. 
 
IF POWERED BY BATTERIES IT IS RECOMMENDED THAT ONLY STANDARD ALKILINE BATTERIES ARE USED. "SUPER HEAVY DUTY" 
BATTERIES NOT UTILIZAING ALKILINE CHEMISTRY CANNOT PROVIDE PERFORMANCE NECCESARY FOR READINGS AND SENSOR 
ACCURACY MAY BE DEMINISHED AS A RESULT.  

Package Contents

Please verify the contents of your package before using the product.  
 
1‐ Sampler with Boot 
1‐ 6' USB Cable 
1‐ Power supply  
1‐ 10' length 1/8 tubing 
2‐ 1/8" Barb tubing bulkhead fitting 
2‐ 1/8" Barb‐10‐32 fittings 
2‐ Inline filters for input side of unit, 1 is spare 
 

Sampler Components

Your device's appearance may vary slightly from the pictured device. 
 
 
 
 
 
 
LCD Display 
 
Displays the current CO2 concentration 
 
 
Removeable Protective Boot 
 
Gently peel boot from top of unit to access 
 
batteries and USB connector for software 
 
 
Power Switch 
 
Turns the unit and the sampling pump ON and OFF  
 
Barb Connections 
Connect provided tubing to these connections for 
closed loop sampling. 
 
131 Business Center, Suite A‐3, Ormond Beach, FL 32174 
: 877.678.4259  | 
TEL
: sales@co2meter.com | www.co2meter.com 
EMAIL
: 866.422.2356 
FAX

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AQ500 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Co2meter AQ500

  • Page 1: Package Contents

        131 Business Center, Suite A‐3, Ormond Beach, FL 32174  : 877.678.4259  |  : 866.422.2356  : sales@co2meter.com | www.co2meter.com  EMAIL CO2 Sample Draw Device Manual Available in 1%, 30%, and 100% CO2 concentration ranges with optional data logging  Introduction Thank you for your purchase of the AQ500 CO2 Sampler. Please read this manual carefully before using your new sampler to ensure  correct operating procedures are followed. The procedures outlined in this manual are vital to the accuracy, longevity, and correct  usage of this product.   Warnings PAY SPECIAL ATTENTION TO BATTERY POLARITY WHEN INSERTING INTO UNIT. INSERTING THE BATTERIES BACKWARDS MAY CAUSE  PERMANENT DAMAGE TO THE UNIT. ONLY USE THE PROVIDED POWER SUPPLY WITH THE UNIT.    IF POWERED BY BATTERIES IT IS RECOMMENDED THAT ONLY STANDARD ALKILINE BATTERIES ARE USED. “SUPER HEAVY DUTY”  BATTERIES NOT UTILIZAING ALKILINE CHEMISTRY CANNOT PROVIDE PERFORMANCE NECCESARY FOR READINGS AND SENSOR  ACCURACY MAY BE DEMINISHED AS A RESULT.   Package Contents Please verify the contents of your package before using the product.      1‐ Sampler with Boot   1‐ 6’ USB Cable   1‐ Power supply    1‐ 10’ length 1/8 tubing   2‐ 1/8” Barb tubing bulkhead fitting   2‐ 1/8” Barb‐10‐32 fittings ...
  • Page 2: Specifications

        131 Business Center, Suite A‐3, Ormond Beach, FL 32174  : 877.678.4259  |  : 866.422.2356  : sales@co2meter.com | www.co2meter.com  EMAIL Minimum System Requirements To utilize our software your Windows‐PC must meet the following requirements:    Windows XP SP3 or higher   Microsoft .Net Framework 3.5 SP1   Pentium 4 2.4Ghz or higher processor   1GB of RAM   Sufficient disk space for logs and application files (20MB minimum, 200MB+ recommended)  Software is compatible with 64‐bit operating systems and is fully tested under Windows 7.   Specifications  Sensor type: non‐dispersive infrared (NDIR)   Measurement Range:   K33 ELG: 0‐10,000 ppm (1%)  K33 BLG: 0‐300,000ppm (30%)  C20 0‐1,000,000ppm (100%)    Sampling Method: Tube sampling   Sensor Life Expectancy:  > 15 years   Maintenance Interval: No maintenance required   Size: 5.75×5.6×1.3in (146×91×33mm)   Tubing sizes: 3/32” or 1/8" ...
  • Page 3     131 Business Center, Suite A‐3, Ormond Beach, FL 32174  : 877.678.4259  |  : 866.422.2356  : sales@co2meter.com | www.co2meter.com  EMAIL 0‐100% Data Logger  Insert 4 AA batteries into this unit. In order to data log the unit MUST be connected to the computer with data  logging off, and DAS software started. Once the unit has been connected click on the “Configure Sensor” button in  DAS, set the data logging period and desired pump interval. We recommend leaving the pump interval to the  default 10 second period.     The 0‐100% model has an internal coin cell CR‐2032 3V battery backup for the real‐time clock. This battery is  inserted in the factory for your convenience and should last the lifetime of the product.   3. Connect Sampling Hoses Connect the sampling hoses to the front of the device. Attach the inline filter on the input side to filter any particles out of  the air stream. For proper operation ensure that the pressure coming into the system is approximately at atmospheric  pressure. Differentials between the sampling pressure and atmospheric pressure will result in pressure‐dependent  measuring errors and leakage of sample within the device which may be undesirable for closed systems.   Powering the Unit The unit can be powered with 4 AA batteries or via the USB cable and wall power supply including in the package.  Battery Power Batteries can solely power the device and are needed for portable measurement. The battery compartment is located on  the rear of the unit and it slides to open. Install 4 good quality AA alkaline batteries.  Wall Power The unit can be operated from the provided wall power supply as well. Batteries are required at all times in 0‐1% and 0‐30%  data logging products for the device to maintain its internal clock operation. The USB plug is located on the top of the unit.    When using the wall power supply, the life of the batteries are extended, as the USB wall power functions as the main  power source.    Note: The yellow boot must be removed to access the battery compartment and the USB plug. ...
  • Page 4     131 Business Center, Suite A‐3, Ormond Beach, FL 32174  : 877.678.4259  |  : 866.422.2356  : sales@co2meter.com | www.co2meter.com  EMAIL         CO2: 0.110%   Data Logging Units For data logging units, the red switch on the front panel starts the logging mode. The display will alternate between current CO2  levels and the current logging settings to provide visual indication of the status of the unit. The unit will always be on, as long as  power is supplied to the unit.     To collect data with the desktop application, connect the unit via the USB cable and click the “Collect Realtime” button from the DAS  software.     To collect logs, connect the unit to a personal computer and launch DAS. Click on the “Manage Logs” button to download and save  data from the unit.   Care, Maintenance, and Calibration Our units typically do not require calibration during standard use. Calibration for units can be performed through our desktop  software, but generally will not be required.    As long as the correct operating procedure is followed, the unit will continue to operate maintenance free. Ensure that the unit does  not become wet, is not exposed to harsh conditions including extreme heat or extreme cold, and is not exposed to liquids like  solvents. Failure to follow these instructions can result in reduced accuracy and even damage the unit. Ensure that at all times only  air is drawn through the sample inlet. Drawing water into the sensor chamber can cause permanent damage.   Warranty This unit comes with a 90 day warranty.    CO2Meter.com warrants our products to be free from defects in materials and workmanship when used for their intended purpose, ...

Table of Contents