R&S
AFQ100A, R&S
1.1.2.4
Web Control
The web control functionality provides remote operation via VNC using a Web browser
(with HTML5). This mode does not require additional installation or activation. The
VNC protocol allows simultaneous operation from several remote devices. The instru-
ment remains locally operable.
The GUI of the R&S AFQ is visible. To perform the settings, you can operate the
instrument as with the manual control. The instrument controls are available via the
front panel simulation. File upload and download between the instrument and the
remote PC is also available.
Starting a Remote Control via the LXI web browser
This section assumes that the instrument and the controller PC are connected in the
LAN.
1. Start a web browser that supports html5 (W3C compliant).
2. Enter the IP address of the R&S AFQ in the browser's address bar.
The R&S AFQ's welcome page is displayed.
3. In the navigation pane, select "Instrument Control" > "Web Control".
Remote access to the instrument requires the password. The default password is
instrument.
4. Enter the password and confirm with the ENTER key.
After the connection is established, the current screen of the R&S AFQ is displayed
in the browser window.
Supplement 1175.6978.02 ─ 04
AFQ100B
LXI (R&S AFQ100B only)
LXI Configuration