Shimadzu GC-201x

4.2.4 Method Setup - GC - Time Events

This tab allows to set the events that should happen during the run based
on the analysis time. This includes events like switching the detector
polarity, range or current (where applicable), setting temperatures to
detector/injector/auxiliary zones etc. To delete a row, simply delete the
value in the Time column on that row.
Note:
This tab uses the same functions as the
described together.
The table should be filled with the Time of the event in the first column (in
minutes, decimal format - 2 minutes 45 seconds should be set as 2.75),
the event type in the Event column, event value in the Value column and
the Line selected as set on the
tab. Possible events are:
Det 1(..4) Polarity
Sets the polarity of the detector (if the polarity for the given detector has
any sense). Polarity + corresponds to the value 1, polarity - corresponds to
the value 2.
Det 1(..4) Range
Changes the detector range (as set before the analysis on the
during the course of the analysis. Range x1 corresponds to value 0, range
x10 -1 to value 1, x10 -2 to value 2, x10 -3 to value 3 and x10 -4 to value 4.
Fig 12: Method Setup - GC - Time Events
Pre Run
Shimadzu GC-201x Setup - Line Config
- 15 -
4 Using the control module
tab, so these are
DET X
tab)