4 Using the control module

4.2 Method Setup - GC

4.2.1 Method Setup - GC - Oven/Zones

Oven Max
Sets the maximal allowed temperature for the column oven. Possible
values are: 0 - 470.
Note:
Please note that the maximum value may differ for certain GC models -
listed values for it range between 350 and 470 °C.
Equilibration Time
Time that the GC needs to reach the temperature equilibrium after the
temperature program gets to the set initial temperature. This time is used
to distribute the heat evenly in the oven.
Gradient Table
Sets the Heat Rate, desired temperature (Final Temp ) and time to hold
this temperature (Hold Time). The Total Time column will be calculated
automatically. In other words, for isothermal part of the temperature
gradient, the temperature set in Final Temp will be kept for the interval set
in Hold Time . To delete a row, simply delete the value in the Heat Rate
column on that row.
Temperature Zones
Sets the temperature and maximum temperature of the zones that have
been set in the
details see also chapter Shimadzu GC-201x Setup on pg 24).
Fig 6: Method Setup - GC - Oven/Zones
Shimadzu GC-201x Setup - Heater Ports
- 8 -
Clarity Controls
dialog (for more