Calibration - Health O Meter 2500KL User Instructions

Digital patient wheelchair scales
Hide thumbs Also See for 2500KL:
Table of Contents

Advertisement

CALIBRATION

Calibration
These scales have been factory calibrated, and does not require calibration prior to use.
For maintenance, the scale can be calibrated. Calibration of your 2500KL or 2600KL scale is
performed using kilograms (kg) or pounds (lb), according to the units of measure used upon
entering into the calibration path. The following calibration assumes the scale is set to pounds,
and 400 lb (the default weight) is employed as the calibration weight.
Scale Calibration
1. Make sure the display is turned off, then press and hold the
HOLD/RELEASE button while pressing the ON/OFF button.
2. Press the ENTER button.
3. Make sure there is nothing on the scale, then press the ENTER
button again.
4. Use the ► and ▲ buttons to set the calibration weight to a
value other than the default weight. The calibration weight should
be as high as possible, with 400 lb as a minimum calibration
weight.
5. Press the ENTER button.
6. Load the scale with the required weight.
7. Without touching or shaking the scale, allow the display to
stabilize, then press the ENTER button.
8. If the calibration is successful, the display will show SUCC and
turn off. If the calibration is not successful, the display will show
Err1 and turn off. Contact customer service for an Err1 display
(+1 800-638-3722).
.
Operator Action
26
Display
"U 01" followed by a 5 digit
'idle scale count' number.
"CAL".
"Clear" followed by
"0400.0LB".
Shows "400.0" or entered
calibration weight.
Alternating "Load" followed
by "400.0LB".
Display increments showing
raw weight data.
"SUCC" or "Err1".
-------

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2600kl

Table of Contents