The following procedure is intended to be used for inspection and periodic calibration checks to confirm that the
7B92A is operating within acceptable limits. This procedure is concerned with those portions of the instrument calibration
that are essential to measurement accuracy and correct operation. Removal of the side covers is not necessary to
perform this procedure in that all checks are made from the front panel.
OUTLINE FOR PART I-
PERFORMANCE CHECK
A.
Trigger Sensitivity
1.
Check Main Triggering Level Range
2.
Check Delayed Triggering Level Range
3.
Check Low-Frequency Triggering
4.
Check AC LF Reject Operation
5.
Check 20 MHz Triggering
6.
Check 500 MHz Triggering
7.
Check HF Sync Triggering
8.
Check Line Triggering
9.
Check Single Sweep Operation
B.
Horizontal System
1.
Set Basic Sweep Calibration
2.
Check Trace Separation Range
(SN B059999-below)
3.
Check Trace Separation Range and
ALT OFF Function (SN B060000-above)
4.
Check Sweep Timing
PART I-PERFORMANCE CHECK
Page
5-5
5-6
5-6
5-7
5-7
5-8
5-9
5-9
5-9
compartment of the oscilloscope.
compartment of the oscilloscope.
5-10
display the vertical unit and the horizontal mode switch to
5-10
display the horizontal unit.
5-10
counterclockwise and set the trigger source switches to
5-10
vertical mode.
minutes warm up before beginning the procedure.
5-4
Performance Check/Calibration-7B92A
5.
Check Sweep Linearity
6.
Check VARIABLE Time/Division Range
7.
Check Differential Delay Time Accuracy
8.
Check Fast Delay Time Accuracy
9.
Check Delay Jitter
PRELIMINARY PROCEDURE
1.
Install the amplifier plug-in unit in a vertical
2.
Install
the
7B92A
3. Set the oscilloscope vertical mode switch to
4.
Set the oscilloscope intensity controls fully
5. Turn on the oscilloscope and allow at least 20
Page
5-12
5-12
5-13
5-14
5-14
into
the
horizontal
@