12.1 Menu mode operation table

Menu mode allows the user to perform calibration and configure parameters such as full scale range, calibration gas level, calibration interval, inhibit current & timeout and set a password. To activate menu mode hold the
magnet over the MENU/ENTER switch for at least 5 seconds and then remove. Menu mode can be password protected to prevent unauthorized person from changing parameters. Initially the password is set to '000' meaning
it is disabled. While in menu mode, the output current of the detector is inhibited to prevent false alarms.
Use the table below to help navigate the menus and make configuration changes. The menus are shown in the left hand column. Use ▲ ▼ to select the required menu and '' to enter. Follow the information and instructions in
the table from left to right from the required menu.
▲ ▼
OK
▲▼
GAS READING, 'ZEro' AND FLASHING ' '
ICON DISPLAYED. APPLY ZERO GAS AND
SEt CAL
1
USE '' WHEN READING IS STABLE. 'ZEro
PASS' DISPLAYED IF OK, FAULT CODE
AND RETURNS TO MENU IF NOT. '
'rAn9' DISPLAYED & CURRENT RANGE
FLASHES. USE ▲▼ TO SELECT
SEt rAn9
DIFFERENT RANGE (SEE SECTION 8.1
FOR AVAILABLE RANGES).
'Curr' DISPLAYED WITH '
CURRENT INHIBIT mA VALUE FLASHES.
ConF
USE▲▼ TO SELECT NEW VALUE (3 OR
Inhb
4mA TOXICS, OR 3 OR 17.4mA OXYGEN
VERSION).
'PASS' DISPLAYED WITH FIRST DIGIT
OF THE PASSCODE FLASHING. USE ▲
SEt
TO SELECT 1ST DIGIT OF CURRENT
PASS
PASSWORD. USE ▼ TO MOVE TO NEXT
DIGIT AND SET REST OF PASSCODE.
Int' DISPLAYED WITH CURRENT
CAL
CALIBRATION INTERVAL FLASHING.
Int
2
USE ▲▼ TO CHANGE INTERVAL.
CURRENT GAS CONCENTRATION DISPLAYED
bumP
WITH OUTPUT INHIBITED ' ' ICON FLASHING.
tESt
APPLY BUMP TEST GAS AND CHECK PEAK
READING ON DISPLAY.
Curr' DISPLAYED WITH '
ForC
FORCE CURRENT '4.00' FLASHES. USE ▲▼
Curr
TO CHANGE TO REQUIRED MA LEVEL.
'SENS' DISPLAYED. REMOVE SENSOR. FIT
NEW SENSOR. DISPLAY SHOWS 'LOAd
CHAn
SENS'. 'SAmE GAS' DISPLAYED ('NEW GAS'
SEnS
3
IF DIFFERENT GAS TYPE SENSOR FITTED)
THEN 'WArm' COUNT DOWN.
QUITS MENU MODE AND RETURNS TO
9u It
MONITORING MODE.
1 Refer to section 8.2 for detailed zero and span calibration procedures. A detector fitted with an oxygen sensor will skip the zeroing procedure. Re-calibrate the detector if left un-powered for periods in excess of 24 hours.
2 The calibration due warning counter is automatically reset after a successful calibration.
3 To exit the Change Sensor Menu without changing the sensor, use ''.
S3KMKIIMAN_MAN0878_Issue 5_07/2013
OK
'SPAN YES' DISPLAYED TO ASK IF
YOU NOW WANT TO PERFORM SPAN
CALIBRATION. USE '' TO PROCEED OR
USE ▲▼ TO SELECT 'No' AND RETURN
TO MENU MODE.
NEW RANGE DISPLAYED. 'YES' FLASHES.
USE '' TO CONFIRM (OR ▲▼ TO
SELECT 'no' AND RETURN TO STEP 1).
' ICON.
'tImE' DISPLAYED WITH '
CURRENT INHIBIT TIMEOUT PERIOD
(MINUTES) FLASHES. USE ▲▼ TO SET
NEW TIMEOUT. (IF SET TO 000 OUTPUT
IS PERMANENTLY INHIBITED).
NEW PASSCODE DISPLAYED WITH 'YES'
FLASHING. USE '' TO CONFIRM (OR ▲▼
TO SELECT 'No' AND RETURN TO SELECT
NEW VALUE).
NEW INTERVAL DISPLAYED WITH 'YES'
FLASHING. USE '' TO CONFIRM (OR ▲▼
TO SELECT 'no' AND RETURN TO SELECT
NEW INTERVAL).
REMOVE GAS. FALLING REAL TIME GAS
READING DISPLAYED.
' ICON. DEFAULT
UNIT RETURNS TO MENU MODE
UNIT RETURNS TO MENU MODE
▲ ▼
OK
CURRENT SPAN CALIBRATION
FLASHING. USE ▲▼ TO CHANGE AND ''
TO CONFIRM. NEW SPAN CALIBRATION
POINT DISPLAYED WITH 'YES' FLASHING.
USE '' TO CONFIRM (OR ▲▼ TO
SELECT 'No' AND RETURN TO SELECT
UNIT RETURNS TO MENU MODE.
' ICON.
NEW TIMEOUT DISPLAYED. 'YES'
FLASHES. USE '' TO CONFIRM (OR ▲▼
TO SELECT 'no' AND RETURN TO SELECT
UNIT RETURNS TO MENU MODE
'duE' DISPLAYED WITH 'No', 'LCD' OR
'both' FLASHING. USE ▲▼ TO SELECT
REQUIRED CAL DUE WARNING OUTPUT.
UNIT RETURNS TO MENU MODE
25
▲▼
OK
GAS READING, 'SPAn' AND FLASHING
' ' ICON DISPLAYED. APPLY SPAN GAS AND
USE '' WHEN READING STABLE. IF OK
'PASS' & 'PurG' DISPLAYED (IF FAIL 'SPAN
FAIL' DISPLAYED AND RETURNS TO MENU).
REMOVE SPAN GAS. WHEN READING <50%
OF SPAN POINT, COUNTDOWN BEGINS &
NEW VALUE).
UNIT RETURNS TO MENU MODE.
UNIT RETURNS TO MENU MODE
NEW VALUE).
UNIT RETURNS TO MENU MODE
▲▼