SERIES 3000 MKII TECHNICAL MANUAL

8.1 Default configuration

Series 3000 MkII detectors are supplied with a default configuration as shown below. The settings for
full scale range, calibration gas level, calibration interval, inhibit current & timeout and access password
can be changed to suit individual applications.
Detector 4-20mA output signal
Fault / Under-range
Inhibit
Zero signal
Full scale
Overrange
Other settings
Inhibit time out
Calibration Interval & Display / Output. Cal Interval 180 days (selectable 30-360). Three options for overdue
Password
Gas
Selectable Full
Scale Range
Oxygen
25.0%/Vol only
Hydrogen Sulfide
10.0 to 50.0ppm
Hydrogen Sulfide
50 to 500ppm
Carbon Monoxide
100 to 1,000ppm 300ppm
Sulfur Dioxide
5.0 to 20.0ppm
Ammonia
50 to 200ppm
Ammonia
200 to 1,000ppm 1,000ppm
Chlorine
5.0 to 20.0ppm
Chlorine Dioxide
1.00ppm only
Nitric Oxide
100ppm only
Nitrogen Dioxide
5.0 to 50.0ppm
Hydrogen
1,000ppm only
Hydrogen
9,999ppm only
Hydrogen Chloride 10.0 to 20.0ppm
Hydrogen Cyanide
30.0ppm only
Hydrogen Fluoride
12.0ppm only
Ozone
0.4ppm only
Phosphine
1.2ppm only
Notes:
1. Default ranges have been selected based upon recommended exposure limits set around the world.
2. Where ranges significantly above the recommended defaults are needed select a gas cal point
near to the required alarm point or between 30 and 70% of full scale. Consideration must be given
to providing adequate ventilation of the calibration gas and ensure there is no exposure of the user
to toxic levels of gas.
For details of how to change the configuration for full scale range, calibration gas level, calibration
interval, inhibit current & timeout and access password refer to section 12.
3mA
3mA (default). Selectable 4mA (toxic), 17.4mA (oxygen)
4mA
20mA
22mA
0 (no time out). Selectable 0 to 240 minutes (5 min steps). If 0 is
selected, output current is permanently inhibited.
warning indication (selectable off, via display only or via display and
3mA output).
000 (password disabled). If changed then password is activated.
Change back to 000 to de-activate.
Default Range
Steps
25.0%VOL
n/a
15.0ppm
0.1ppm
100ppm
10ppm
100ppm
15.0ppm
5.0ppm
200ppm
50ppm
50ppm
5.0ppm
5.0ppm
1.00ppm
n/a
100ppm
n/a
10ppm
5.0ppm
1,000ppm
n/a
9,999ppm
n/a
10ppm
1.0ppm
30.0ppm
n/a
12.0ppm
n/a
0.4ppm
n/a
1.2ppm
n/a
17
S3KMKIIMAN_MAN0878_Issue 6_01/2014
Selectable Cal
Default Cal
Gas Range
Point
20.9%/Vol (Fixed) 20.9%/Vol
10ppm
50ppm
100ppm
5.0ppm
100ppm
300ppm
2.0ppm
0.5ppm
30 to 70% of
selected full
50
scale range
5ppm
500ppm
5000ppm
5.0ppm
10.0ppm
5.0ppm
0.2ppm
0.5ppm