Installation Instructions

Place on a sturdy and level surface in a safe, dry place, and connect with a
buffer filled electrophoresis system.

Operation Instructions

1. Place the power supply on a sturdy and level surface in a safe, dry place,
away from laboratory traffic.
2. Ensure that the AC power switch is OFF, and then plug the three-pronged
power cord into a grounded three-prong AC outlet of the appropriate voltage
(110V to 240V as indicated on the rating sticker near the AC cord on the
back of the unit).
3. Turn the AC power ON by the switch on the rear of the machine.
4. Connect the DC output jacks from the electrophoresis unit; insert the red
lead (+) into the red output jack, and the black lead (-) into the black output
jack.
5. Press
Key, and then press
suitable voltage.
6. Press
Key, and then press
suitable current.
The power supply will then run in constant voltage at the voltage set or
constant current at the current set, depending on which setting is
exceeded first. If the current is too low at the voltage set, the voltage
setting should be increased. If the voltage is too low at the current set,
the current should be increased.
Key or
Key or
11
Key to select the
Key to select the