Satelec Mini LED Autofocus User Manual page 13

Dental curing light
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
an electromagnetic environment measurement must be made. If the measured radiofrequency field strength in
the immediate environment where the product is used exceeds the compliance level specified above, the
performance of the product must be tested to verify whether it conforms to the specifications. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.
12.4 - Recommended separation distances
The Mini L.E.D. AutoFocus is intended for use in an electromagnetic environment in which radiated radiofrequency
disturbances are controlled.
The Mini L.E.D. AutoFocus user and/or installer can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile radiofrequency communications equipment (emitters) and the Mini
L.E.D. AutoFocus, according to the maximum output power of the equipment, as recommended in the table below.
Rated max. power
of the emitter (W)
0.01
0.1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the emitter, where P is the max. power rating of the emitter in watts (W) according
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
12.5 - Cable lengths
Cables and
Maximum
accessories
length
charger power
<3m
cord
Separation distance in meters (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
the manufacturer.
RF emission, CISPR 1 - Class B
Immunity to magnetic fields: IEC61000-4-8.
Immunity to electrostatic discharge: IEC61000-4-2
Immunity to electrical fast transients/bursts: IEC61000-4-4
Immunity to surges: IEC61000-4-5
Immunity to voltage dips, short interruptions and voltage variations:
IEC61000-4-11
Immunity to conducted disturbances induced by radiofrequency fields: IEC61000-4-6
Immunity to radiated radiofrequency electromagnetic fields: IEC61000-4-3
80 MHz to 800 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
Complies with:
13
800 MHz to 2.5 GHz
d = 2.3
0.23 m
0.73 m
2.3 m
7.3 m
23 m