12.2 - Magnetic and electromagnetic immunity
The
Mini L.E.D. AutoFocus
below.
The user and/or installer must ensure that the
environment.
Immunity test
Magnetic field at 50Hz.
3A/m
IEC61000-4-8
Electrostatic discharge
± 6KV contact
(ESD)
± 8KV air
IEC 61000-4-2.
Electrical fast transients
± 2KV for power supply
IEC 61000-4-4.
lines
Surges
± 1KV differential mode
IEC 61000-4-5.
± 2KV common mode
<5% UT
for 0.5 cycles.
Voltage dips, short
40% U
interruptions and voltage
5 cycles
variations
70% U
IEC 61000-4-11.
25 cycles
<5% U
250 cycles
12.3 - Electromagnetic immunity / mobile radiofrequency equipment
The Mini L.E.D. AutoFocus is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the Mini L.E.D. AutoFocus (including its cables) at a distance
less than that recommended and calculated according to the frequency and power of the emitter.
Conducted disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic environment measurement (a), must be
less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile telephones
(cellular/cordless), mobile radios, amateur radios, AM/FM radio broadcasts and TV broadcasts cannot be
determined exactly by theory. To assess the electromagnetic environment due to fixed radiofrequency emitters,
is intended for use in the electromagnetic environment specified in the table
IEC 60601 test level
3A/m
± 6KV contact
± 8KV air
± 2KV for power supply
lines
± 1KV differential mode
± 2KV common mode
(>95% dip in U
)
<5% U
T
T
0.5 cycles.
(60% dip in U
) for
40% U
T
T
cycles
(30% dip in U
) for
70% U
T
T
25 cycles
(>95% dip in U
) for
<5% U
T
T
250 cycles
IEC 60601 test level
3 V/m
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Mini L.E.D. AutoFocus
Compliance level
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Power quality should be that of a typical commercial or hospital
environment.
Power quality should be that of a typical commercial or hospital
environment.
(>95% dip in U
) for
T
T
Power quality should be that of a typical commercial or hospital
(60% dip in U
) for 5
T
T
environment.
If the use of the Mini L.E.D. AutoFocus requires continued
(30% dip in U
) for
T
T
operation during a power cut, it is recommended that the
product be powered from a separate power supply (UPS, etc.).
(>95% dip in U
) for
T
T
aCompliance level
Recommended separation distance:
3 V/m
d = 1.2
d = 1.2
d = 2.3
3 V/m
Where P is the maximum power rating of the emitter in watts (W)
according to the manufacturer's specifications and d is the
recommended minimum separation distance in meters (m).
12
is used in such an electromagnetic
Electromagnetic environment - guidance
Electromagnetic environment - guidance
80MHz to 800MHz.
800MHz to 2.5GHz.