OPERATION
MODEL 148 NANOVOLTKETER
d.
RANGE Switch.
The RANGE Switch
selects
the full-scale
meter sensitivity
(either
microvolts
or millivolts)
for one of nine ranges,
from 0.01 to 100.
e.
FUNCTION Switch.
The FUNCTION Switch selects
the function
-
MICROVOLTS
or MILLIVOLTS -
which is to be measured.
f.
ZERO SUPPRESS Controls.
Two controls
determine
the amount of zero suppression.
1.
The COARSE Control
disconnects
the suppression
circuit
(in OFF position)
or selects
one of four
suppression
voltages
in decade steps.
Refer to Table 3.
2.
The FINE Control
is a continuously
variable
adjustment
for the suppression
voltage
set by the COARSE Control.
It adjusts
the range between the positive
and negative
values
of the maximum voltage
set by the COARSE Control.
g.
INPUT Receptacle.
The INPUT Receptacle
is of a special
low-thermal
design.
Use only the Models 1481, 1482 and 1486 for mating connectors.
WER
5103:
RD
FIGURE 3.
Model 148 Rear Panel Controls
and Connections.
Circuit
designations
refer
to Replaceable
Parts
List
and schematic
diagrams.
2-2.
REAR PANEL CONTROLS AND CONNECTIONS.
a.
Line Voltage
Switch.
The screwdriver-operated
slide
switch
sets the
Model 148 for 117 or 234~volt
ac power lines.
b.
Fuse.
1.
For 117~volt
operation,
use a 3 AG or MDL Slow-Blow
l/8-ampere
fuse.
2.
For 234volt
operation,
use only a MDL Slow-Blow
l/16-ampere
fuse.
c.
Power Cord.
The 3-wire
power cord with
the NEMA approved
3-prong
plug
provides
a ground connection
for the cabinet.
An adapter
for operation
from
2-terminal
outlets
is provided.
6
0464