6.7 Flow verification

Flow verification routine is used to check if internal flow-meters need calibration. No
instrument settings are changed during the flow verification routine.
For flow verification you need a flow calibration pad and an external flow-meter, which you
connect to the Aethalometer inlet.
To start flow verification select "Flow verification" button on Operation/General screen. When
the procedure is finished you will get the table with measured flows, reported at the flow
reporting standard selected for the instrument. Flow verification report:
Manual flow verification report:
02 Oct 2014 14:41:54
External flowmeter measurement:
P
T
Fin
101325 25
886
101325 25
3000
101325 25
5000
Flow verification results:
Flow reporting standard: AMCA 101325 Pa 21.11 °C
Fin
F1
(%)
874
851
(97)
2961
2921
(99)
4935
4880
(99)
If measured flows differ for more than 10 % you need to recalibrate the flow.
____________________________________________________________________________________________________________
User's manual – Ver. 1.53
Fc
(%)
865
(99)
2933
(99)
4880
(99)
July 2015
70/75