6.5 Manual clean air test

Clean air test is conducted using the built-in filter to determine the performance of
Aethalometer under the standard flow condition.
To perform the clean air test, select Operation/General/Clean air test. Test runs for 20
minutes at 5 lpm and 1 s time resolution. Clean air test is marked by status 2048. After the
test a report is generated:
Clean air test report.
Serialnumber: AE33-S02-00232
Date and time: 01 Dec 2014 13:41:03
Duration: 00:20:00, Timebase: 1 sec, Flow: 5000 mlpm
AverageBC
Spot1 Spot2 Spot1 Spot2
Ch1
92
160
Ch2
38
76
Ch3
37
89
Ch4
46
128
Ch5
47
108
Ch6
24
94
Ch7
32
107
Result of clean air test is acceptable.
Interpretation of the results:
- Average BC values should be close to zero if the Aethalometer is warmed up (was
measuring for a few hours just before conducting the test). It is possible to observe a
short transient due to filter compression.
- Point to point variation of BC (PPBC) is calculated as an average absolute difference
between the consecutive BC measurements. It depends on the selected timebase, flow
and tested channel. A test criterion for 1 s time-base test at 5 lpm flow is: PPBC61 < 550
ng/m
.
3
____________________________________________________________________________________________________________
User's manual – Ver. 1.53
PPBC
(ng/m3)
193
398
255
592
268
600
307
672
308
668
385
848
437
916
July 2015
68/75