Technical Specification - Elcometer SCM400 Operating Instructions Manual

Salt contamination meter
Table of Contents

Advertisement

9 TECHNICAL SPECIFICATION

Chemical
Measuring range:
Resolution:
Measuring accuracy:
Temperature correction:
Electrical
Electrodes:
Display:
Sampling timer:
Battery:
Timeout:
Working limits
Temperature range:
Maximum relative humidity: 80%
Physical
Dimensions:
Weight:
Case:
Sealing:
Sampling capability
Substrate:
Area required:
Position:
-2
0.1 µg.cm
to 20 µg.cm
sodium chloride) when using recommended
High Purity Sample Papers and water with a
conductivity of <0.25 microsiemens
-2
0.1 µg.cm
(with sodium chloride)
±10% of reading
Calibrated at 25° C
Below 25°C: -1.7% /° C
Above 25° C: +1.7% /°C
Copper
20 digit LCD
2 minutes
9 V Alkaline PP3 (6LR61)
20 seconds
5° C to 40° C
200 mm x 190 mm x 60 mm
1.5 kg
Damage resistant polyurethane
Splash proof
Metallic or non-metallic
110 mm diameter
Any horizontal, vertical, curved or inclined
surface where sample paper can be placed
-2
soluble salt (as
19