9. With test strip still in meter, touch
Sample Tip of test strip to top of
drop of control solution. Allow
drop to be drawn into test strip.
Remove test strip from drop when
meter beeps.
10. Dashes appear across the Display to
show meter is testing.
Note: If meter does not beep and begin testing soon
after drawing up sample, release and discard
test strip. Repeat test with new test strip. If
problem persists, see Troubleshooting.
11. Compare meter result to Control
Test range printed on test strip
vial label for level of control
solution you are using. If result
is in range, System can be used
for testing blood. If result does
not fall within range, repeat test
using a new test strip.
Note: Control Test result shows the
Control Symbol in the Display.
19
Meter testing
Control Solution
Bottle Label
Test Strip
Vial Label
Control Symbol
(Example only. Does
not represent actual
Control Test ranges.)