Precautions And Adverse Reactions - Zynex Medical InWave User Manual

Multiple-mode stimulator
Table of Contents

Advertisement

Warnings (Cont.)
Maximum RMS voltage (500 Ω load) = 2 V
Maximum RMS current (500 Ω load) = 4 mA
Maximum output power (500 Ω load) = 8 mW
Maximum RMS voltage (1 KΩ load) = 4 V
Maximum RMS current (1 KΩ load) = 4 mA
Maximum output power (1 KΩ load) = 16 mW
Maximum RMS voltage (2 KΩ load) = 5 V
Maximum RMS current (2 KΩ load) = 2.5 mA
Maximum output power (2 KΩ load) = 12.5 mW
A yellow LED indicator is provided next to the probe connector to show
that the unit is delivering energy for any non-zero value of stimulation
output (1 to 80 mA).
The long-term effects of chronic electrical stimulation are unknown.
Safety has not been established for the use of the InWave during
pregnancy.
Discontinue use of this device in the event of pain or bleeding.
Do not use this device in water.
Warning labels are legible from a distance of 30 cm or less, at a 90° angle
or less, and at 500 lux or greater luminous intensity.
Precautions
Stimulation settings should be based on the guidance of the prescribing
practitioner.
This device should be kept out of the reach of children.
This device should only be used with the vaginal probe(s)
recommended for use by the manufacturer.
This device should not be used while driving, operating machinery, or
during any activity in which involuntary muscle contractions may put the
user at risk of injury.
Patients with implanted electronic devices, e.g., a cardiac pacemaker,
should not use this device unless advised by a physician.
Simultaneous connection of a patient to high frequency surgical equipment
may result in burns at the site of the vaginal probe and possible damage to
the stimulator.
Operations within close proximity (within one meter) of shortwave or
microwave therapy equipment may produce instability in the stimulation
output.
Warnings
Urge UI Mode
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents