Stress Urinary Incontinence (UI) Mode
Program
5/10
The pulse width is 200 µs, and the frequency is 50 Hz. The output signal
ramps up to the stimulation level set by the user in one second. The output
stays on for five seconds, then ramps down to a zero level in one second.
The output stays off for ten seconds. The cycle repeats for the duration of
the treatment time. The default treatment time is 15 minutes.
3/10
Similar to 5/10 program except the on time is three seconds.
5/5
Similar to 5/10 program except the off time is five seconds.
Urge UI Mode
Program
10/3
The pulse width is 200 µs, and the frequency is 12.5 Hz. The output signal
ramps up to the stimulation level set by the user in one second. The output
stays on for ten seconds, then ramps down to a zero level in 0.5 second.
The output stays off for three seconds. The cycle repeats for the duration
of the treatment time. The default treatment time is 15 minutes.
6/3
Similar to 10/3 program except the on time is only six seconds.
10/10
Similar to 10/3 mode except the off time is ten seconds.
Mixed UI Mode
Program
10:10
A 5/10 Stress program runs for ten minutes, followed by a ten minute 10/3
Urge program. This cycle repeats for the duration of the treatment time.
The default treatment time is twenty minutes.
5:5
Similar to 10/10 program except the 5/10 Stress program only runs for five
minutes and the 10/3 Urge program only runs for five minutes.
2:2
Similar to 10/10 program except the 5/10 Stress program only runs for two
minutes and the 10/3 Urge program only runs for two minutes..
The InWave should be used daily for 15 to 20 minutes per session, for 1 to 3 months, or
as directed by a physician.