Type PJ 9080-023.3, 110/230 Volts A.C., 50/60 Hz, IP 20 Weight: approx. 6,4 kg Supply schedule: Special shipping carton, containing: 1 VITA VACUMAT 500 furnace 1 control panel 1 firing tray 1 mains power lead 1 pair of furnace tweezers...
Place the firing tray onto the lift support plate. Notice: Never operate the VITA VACUMAT 500 without the firing tray on the lift support plate. It may cause damage to the furnace. Press the "Start" key. The firing tray ascends into the firing chamber, and the temperature starts risinguntil it has reached the factory-set starting temperature of 500°C.
VITA – Vacumat 500 Control panel Block A Block B Block C Block D Block E Block F Block G Description (Block A) Firing chamber temperature 20°C - 1200°C 1. Program No. 0 Close withaut heating 2. Program No. 1 - 200 Firing programs 3.
VITA – Vacumat 500 Description Heating-up (Block C) Heating-UP LED 1. flashes when Input-Mode is active 2. „ON“ indicates a firing program has been selected 3. flashes when HEATING-UP is in program Display shows: temperature rising rate in 20°C - 130°C Actives Input-Mode for temperature rising rate Description Firing temperature &...
VITA – Vacumat 500 Description Slow / Fast cooling (Block F) COOLING LED 1. flashes when Inpu-Mode COOLING is a active 2. „ON“ when COOLING is a step of the selected program 3. flashes when COOLING ist active Display shows: COOLING temperature 300°C - 1000°C...
VITA – Vacumat 500 Operating Block B – Pre-drying temperature & Pre-drying time Select program using keyboard and press Enter-Key '#'' Activate Input-Mode with When Input-Mode is active, PRE-DRYING Key PRE-DRYING key will switch between input of PRE-DRYING time and PRE-DRYING temperature.
VITA – Vacumat 500 Operating Block C - Heating up Select program using keyboard and press Enter-Key '#' Activate Input-Mode with HEATING-UP key flashing HEATING-UP LED LCD shows ??? Press Change temperature Enter-Key'#' rising rate? Valid Input: 20°C - 120°C/min...
VITA – Vacumat 500 4.10 Operating Block D - Firing temperature & End temperature select program using keyboard and press Enter-Key '#' Activate Input-Mode with END temperature key When Input-Mode is active, flashing END temperature LED END temperature key will switch LED min:sec HOLD time "ON"...
VITA – Vacumat 500 4.11 Operating Block E - Vacuum parameters Select program using keyboard and press Enter-Key '#' Activate Input-Mode with VACUUM key flashing VACUUM LED LCD shows VACUUM time in min:sec (see *) Input-Mode may be interrupted at any time by pressing the...
VITA – Vacumat 500 4.12 Operating Block F - Slow cooling Select program using keyboard and press Enter-Key '#' Activate Input-Mode with COOLING key flashing COOLING LED Input-Mode may be LCD shows COOLING temp. °C interrupted at any time by pressing the 'STOP' key.
Protection against Power failure The VITA VACUMAT 500 is protected against power failure. This backup device is activated immediately in the event of a mains power failure during a firing or stand-by program. If the power failure lasts for less than 20 seconds the program continues.
VITA – Vacumat 500 Utilities All programs listed may be selected by entering the respective program number using the keyboard (BlockH). The selection must be confirmed by pressing the "#"-Key. The programs may only be selected if no firing program is active. To stop or cancel a program, press the "Stop"-Key...
Changing the muffle • Only to be carried out by a qualified service technician! • Contact your VITA dealer or authorized service personnel! Switch off furnace and disconnect from the mains power supply! 1. Unscrew the 2 screws on the rear panel and remove it.
VITA – Vacumat 500 Error Messages / Service The VITA - VACUMAT 500 furnace is equipped with a self - checking system which detects and indicates possible errors. The following error messages may occur: Err 1 the required i.e. the preselected temperature has been exceeded by 80 °C.
Need help?
Do you have a question about the Vacumat 500 and is the answer not in the manual?