Care and Maintenance
CAUTION: Do not attempt to use a battery
that was not supplied by BHM Medical.
BHM Medical
batteries
designed
for
BHM Medical
systems.
Attempting
unauthorized
battery
damage the lift and/or the charger.
Verification of the Charger's Power
Source
If the light does not illuminate when there are
batteries correctly installed in the Voyager Duo, try
the following:
1) Make sure that the power cord is correctly
plugged into the charger and in the wall AC
outlet, and that the green light on the clip-on
charging station is on.
2) Make sure that there is contact between the
contact blades of the lift and the contact plates of
the charging station.
3) Check the power of the AC outlet on the wall.
4) If the charger's green light does not light up,
contact your local BHM Medical representative
for assistance.
Sling Inspection and
Care
For maximum resident safety and hygiene, read the
instructions found in the Loop Sling Instructions for
Use that comes with the sling.

Annual Inspection

The Voyager Duo and its accessories must be
inspected annually by a qualified technician.
WARNING:
The
Voyager
accessories must be serviced every 12
months as a minimum requirement. Do not
attempt to do the inspection unless you are
certified to do so.

Maintenance Requirements

The Voyager Duo is equipped with an electronic
monitor that causes a red light to flash when a
are
specially
maintenance inspection is necessary. Arranging for
charging
scheduled inspections ensures the durability of the
to
use
an
unit and the security of the resident and user.
may
seriously
Duo
and
47