Charger Specification; Maintenance; Daily Check List; Inspection And Cleaning - BHM Medical VOYAGER 550 Technical Manual

Battery and electric version 2 and 4 function ceiling lift system
Table of Contents

Advertisement

the
button on the handset. The lift cannot
return to charger with a 50-pound load.
For the 2 function, pull the lift gently over to the
charger and ensure it is properly connected.
DO NOT SLIDE THE LIFT OVER TO
THE CHARGER FORCEFULLY OR
QUICKLY. THE CHARGER MAY
BECOME DAMAGED.
The indicator light will illuminate red to green
depending on the battery drain. If the light
does not go on, check "Troubleshooting"
section in this manual for assistance.
If the battery is low, the light will show red or
orange when you return the lift to the
charger.
The light will gradually turn to
green when fully charged.
Whenever possible leave the lift
on the charger when the lift is not
in use. At minimum, charge the
battery until the light is green
before using it again.
extend the life of the battery.

Charger Specification:

Hospital grade charger cord and plug.
Charger input: 120 Vac, 60 Hz, 0.5 A. (North
America) / 240 Vac, 50 Hz, 0.25 A
(European).

MAINTENANCE

ALWAYS CARRY OUT THE DAILY
CHECKLIST BEFORE EACH LIFT
USE.
Alterations made to the VOYAGER 550 by
someone other than a certified technician
may cause serious injury and will void
warranty.
The VOYAGER 550 and accessories must
be inspected ANNUALLY by a certified
technician in addition to the daily and other
periodic visual checks done by the user
specified in this section.
Preventive maintenance specified in this
manual can prevent accidents and reduce
repair costs.
Note all service or repairs to the VOYAGER
550 or its accessories in the log book at the
end of this manual. Have the document
signed by the certified technician.
Technical-VOYAGER-550 (REV 4)

Daily Check List

THE FOLLOWING PROCEDURES MUST
BE FOLLOWED BEFORE EACH USE.
Has the battery been charged? Park the lift
on the charger whenever the lift is not in use.
Inspect the lift for any damage. If the lift
casing does not look properly aligned, or
there are any cracks or other damage on the
lift, or any parts are missing - DO NOT USE
IT.
have the lift serviced.
Inspect the visible strap for any signs of
wear, frays, loose threads or other damage.
If there is any evidence of damage- DO NOT
USE IT. Contact your local representative to
have the lift serviced.
Inspect the sling for tears, frayed straps or
loose stitching. If the sling has any of the
above damage, DO NOT USE IT. Contact
your local representative to have the sling
replaced or repaired.
This will
Inspect the carry bar for any signs of
cracking or damage.
Ensure the ring and cotter pins that attach
the carry bar to the strap are secure.

Inspection and Cleaning

Clean the Voyager 550 with a soft dry cloth, or a
soft cloth lightly moistened with a mild detergent
solution. Do not use any type of solvent that may
damage the finish.
To ensure a better rolling surface for the trolley
wheels, clean the inside of the track every 4
months. To do so, insert a damp cloth in the
opening and slide it from one end of the track to
the other.

Strap Inspection

This equipment is built with a strap that can
withstand a load of 1364 kg (3000 lbs.).
Nonetheless, if this strap is damaged or shows
signs of wear, the acceptable load on the strap
before rupture can drop rapidly and present a
danger for the patient and/or caregiver.
MAINTENANCE
Contact your local representative to
Do not immerse lift in water.
Always reinstall rail end stoppers
(if they have been removed) after
servicing.
11

Advertisement

Table of Contents
loading

Table of Contents