Turbo V™ Ion Source Settings; Compound Parameters - AB Sciex API 3200 System Hardware Manual

Fully-integrated triple quadrupole mass spectrometer
Table of Contents

Advertisement

Table 1-1 Source/Gas Parameters
Curtain Gas (CUR)
IonSpray Voltage (IS)
Temperature (TEM)
Ion Source Gas (GS1)
Ion Source Gas 2 (GS2)
Turbo V™ Ion Source Settings
Table 1-2
parameters. For more information, see the Turbo V™ Ion Source Operator's
Guide.
Table 1-2 Vertical and Horizontal Parameters
X-axis
Y-axis

Compound Parameters

In general, you can use the preset values for most of the parameters in
For more information about the parameters, see the Analyst
system.
Table 1-3 Compound Parameters
CAD Gas
DP (Declustering Potential)
EP (Entrance Potential)
CXP (Collision Cell Exit
Potential)
CE (Collision Energy)
IE1 (Ion Energy 1) — For
experienced instrument
operators only
IE3 (Ion Energy 3) — For
experienced instrument
operators only
Document Number: D1000092204 B
Parameter
shows the recommended starting values for the X- and Y-axis
Parameter
Parameter
Value
20
5000
700
60
60
Value
5
5
Use the preset value and optimize for your
compound.
Use the preset value and optimize for your
compound.
Use the preset value.
Use the preset value and optimize for your
compound.
Use the preset value and optimize for your
compound.
Do not adjust.
Do not adjust.
Hardware Guide
Table
software Help
®
Value
1-3.
15