Setup - Grundfos Conex DIS-C Installation And Operating Instructions Manual

Measuring instrument of conductivity
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

12.4 Setup

The "Setup" menu contains settings that need to be
made or checked the first time the Conex
used.
1. Press the [Down] button to navigate to the
"Setup" menu.
Alarm
Setup
2. Press the [Right] button to open the "Setup"
menu.
3. Press the [Down] button to navigate to the
submenus.
In the submenus, it is possible to enter settings
for the following values:
temperature correction
control parameters
switch-on delay
analog output
language
temperature coefficient
cell constant C
bus address
cable compensation/zero point
measuring range
averaging on/off.
12.4.1 Temperature correction
Because the temperature sensor is connected using
a 2-conductor method, this can result in deviations in
temperature measurement and display.
These deviations can be corrected within a range of
± 5 °C.
1. Press the [Right] button to open the "Corr. temp."
menu.
Corr. temp.
0.0 °C
2. Press the [Right] button to start value input.
3. Press the [Up] and [Down] buttons to set the
value.
12.4.2 Control parameters
The first time the device is used, the regulatory
functions must be defined.
®
DIS-C is
In the "Contr. param." menu, the following values
can be set:
pulse frequency S1
pulse frequency S2
hysteresis
pulse + pause
minimum impulse.
1. Press the [Right] button to open the
"Contr. param." menu.
2. Press the [Down] button to navigate to the
submenus.
Pulse frequency S1/S2
Note
Note
Note
1. Press the [Right] button to open the "Pulse
freq. S1" or "Pulse freq. S2" menu.
2. Press the [Up] and [Down] buttons to set the
value.
Corr. temp.
Contr. param.
Setting for a limit value: Xp = O.
Setting for an intermittent pulse regulator:
Xp > O, pulse frequency = 0.
Setting for a pulse frequency regulator:
Xp > O, pulse frequency > 0.
Pulse frequency setting = maximum
number of impulses per minute. Set the
appropriate pulse frequency for the
connected control element.
It is possible to make separate control
parameter settings for S1 and S2.
Corr. temp.
Contr. param.
21

Advertisement

Table of Contents
loading

Table of Contents