The manual shall not be handed over, or made ........ 10 Checklist before use available to, any unauthorised third party, wit- Disassembly ..........10 hout a prior written consent from Molift AS. Exciting laws, conventions and agreements protects all documents. No extract of this do- Operation ............11 cumentation can be reproduced, used or trans- Battery care ..........11...
Before using the patient lifter ted. Molift AS will accept no liability for damages Before using a patient lifter, the operator must be resulting from such acts.
All instruction and information in this manual re- Injury to personnel or equipment may occur if garding safety and general precautions must be read, Molift QUICK RAISER is used in an incorrect man- understood and followed carefully. Any disregard to ner and if: these precautions may result in personnel injury or damage to the equipment.
About the product Main components The chassis is the main part of the QUICK RAISER Molift QUICK RAISER is designed to cover the consisting of a steel frame, casters, footplates and need for lift and transfer of patients who are to some emergency stop.
1070 x 610 x 180 mm (LxWxH) Incl. battery: 30 kg Battery: Measurements in illustrations below are in millime- 14,4 V NiCd 1,9 Ah or 14,4 V NiMH , tres. Ah, Molift PowerPac Number of lifts (75 kg, 500 mm up/down): Ca. 40 Page 6 of 0 E Rev. G/ 09/006...
1 V DC linear actuator Weight, total: Measurements in illustrations below are in millime- Incl. battery: approx.. 37 kg Battery: tres. 14,4 V NiCd 1,9 Ah or 14,4 V NiMH , Ah, Molift PowerPac Number of lifts (75 kg, 500 mm up/down): Ca. 40 Page 7 of 0...
1 V DC linear actuator Weight, total: Measurements in illustrations below are in millime- Incl. battery: approx. 38 kg Battery: tres. 14,4 V NiCd 1,9 Ah or 14,4 V NiMH , Ah, Molift PowerPac Number of lifts (75 kg, 500 mm up/down): Ca. 40 Page 8 of 0 E Rev. G/ 09/006...
T-groove at the base of the col- umn and approx. 25mm up. Tighten one of the Molift AS tests all lifters with and without load before shipment. The control is an extra screws slightly to fix the plate. Insert the col-...
Checklist before use • Verify that the lifting arm is properly fastened to the trolley • Verify that the lifting column is properly placed into the chassis - arrow shall be exactly aligned with the edge of the chassis • Verify that the column is properly fastened with bolts tightened Disassembly •...
It’s not possible to overcharge the batteries, but to ensure optimal performance through- Battery care out the lifetime of the battery, we recommend Molift PowerPac system is a battery with a 14,4 V that the battery is allowed to rest for a cou- cell unit that is placed in the battery holder on the ple of minutes after charging to cool down lift. A charger is also supplied to charge the battery before use. It is also important that fully...
Lifting Emergency stop and lowering • Lower the lifting arm to the lowest position All of the Molift patient lifts are equipped with • Alternatively adjust the legs of the QR2 to adapt emergency stop and lowering. The emergency stop...
Page 13
If necessary - help the patient placing the feet correctly. • Place the Molift Easy Raiser sling under the arms and around the back of the patient as far down as possible. Connect the belt and tighten firmly.
• On QR2+ the ropes is attached to the lifting arm Movement bt pulling them firmly from above into the rope The patient should not be standing straight locks as shown on the illustration. up during movement - this can be experi- enced as uncomfortable.
The strap is adjustable Size Molift Easy Raiser Art. no 300003 Art. no 300000 Art. no 300001 Art. no 30000 Art. no 300004 Molift Easy Raiser lifting strap with support Size Molift Easy Raiser with support Waist meas- Art. no 3019003 Recom- Colour urement Art.
Symptom Possible cause/action Maintenance Lifting column is Locking bolts are not prop- wobbly erly secured or lifting column is not properly in place in the Daily maintenance chassis / check if the col- umn is correctly placed in the Detergents must be pH-neutral. Do not use chassis and/or tighten bolts solvents or strong liquids - this may damage Lifter only moves...
If you do not know whom to contact locally, please contact your dealer or Molift AS to get help. Yearly service is done according to procedures in the service manual included with the lifter.
Molift AS hereby declares that this product is awarded up to 5 - five - year extended manufacturer’s warranty. It is very important to our customers and to Molift AS that our products meet all relevant quality and sa- fety requirements according to both customer and authority demands. Molift AS develops and manufac- tures products with high standards of safety, quality and design, and we are continuously in a process of keeping our products in position among the market leaders.
• The Council Directive concerning medical devices (93/42/EØF) and according to this classified as medical equipment class 1 Molift QUICK RAISER is manufactured in conformity with the following national or international standards that also might implement a harmonised standard: • ISO10535 Hoists for the transfer of disabled persons - Requirements and test methods •...
This is the insight that drives us at Molift. The company was started 20 years ago by a man who through his own experience saw the need for lifting and moving patients in an effective and comfortable way - for both patient and carer.
Need help?
Do you have a question about the QUICK RAISER and is the answer not in the manual?
Questions and answers