Salter BPW-9100 Instructions And Guarantee page 10

Fully automatic electronic blood pressure monitor
Table of Contents

Advertisement

11. TROUBLESHOOTING (2)
GB
PROBLEM
LCD shows low battery symbol
LCD shows "Er 0"
LCD shows "Er 1"
LCD shows "Er 1"
LCD shows "Er 1"
LCD shows "Er 4"
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er A"
No response when you press button
or load battery.
NOTICE
01.
Do not drop this monitor or subject it to strong impact.
02.
Avoid high temperature and solarization. Do not immerse the monitor in water as this will result in damage to the
monitor.
03. If this monitor is stored near freezing, allow it to acclimate to room temperature before use.
04.
Do not attempt to disassemble this monitor.
05. If you do not use the monitor for a long time, please remove the batteries.
06. It is recommended the performance should be checked every 2 years or after repair. Please contact the service center.
07. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after moistened with water, diluted disinfectant
alcohol, or diluted detergent.
08. No component can be maintained by user in the monitor. The circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the user's appropriately qualified technical personnel to
repair those parts of equipment which are designated repairably can be supplied.
09. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements or three
years, and the cuff integrity is maintained after 1,000 open–close cycles of the closure.
10. It is recommended the cuff should be disinfected 2 times every week if needed (For example, in hospital or in clinique).
Wipe the inner side (the side contacts skin) of the cuff by a soft cloth squeezed after moistened with Ethyl alcohol
(75-90%), then dry the cuff by airing.
POSSIBLE CAUSE
LCD shows low battery symbol
Pressure system is unstable before
measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff is
too tight during inflation
Pneumatic system leakage or cuff is
too loose during inflation
Cuff pressure above 300mmHg
More than 3 minutes with cuff
pressure above 15 mmHg
EEPROM accessing error
Device parameter checking error
Pressure sensor parameter error
Incorrect operation or strong
electromagnetic interference
10
SOLUTION
Change the batteries
Don't move and try again.
Apply the cuff correctly and try again
Measure again after five minutes. If
the monitor is still abnormal, please
contact the local distributor or the
factory.
Take out batteries for five minutes,
and then reinstall all batteries.