General Troubleshooting - Freestyle MN126-1 Service Manual

Table of Contents

Advertisement

7.0 Troubleshooting

7.1 General Troubleshooting

Before reviewing the Troubleshooting Chart, the following steps may be useful to
isolate any malfunctions:
1. Make sure all filters are clean.
2. Turn the unit on by pressing flow setting [ 2]. A brief, continuous alarm sounds
to indicate that the FreeStyle has been turned on. The compressor will start to
run, you must wait 10 seconds before proceeding to step 3.
3. Activate the test mode by pressing flow settings [1] and [ 3] simultaneously
for five seconds.
4. Make sure that the unit is leak-free by testing all tubing connections and
fittings with leak-testing solution. Protect the circuit board from solution, and
start the leak test at the compressor, following the air flow of the unit to the
oxygen outlet. Repair all leaks by tightening connections and fittings.
5. After five minutes, measure the oxygen concentration.
6. Review the Troubleshooting Chart that follows to isolate and repair any other
malfunction.
FreeStyle Service Manual MN126-1 E
7-1