Clean or replace the air intake / gross particle filter with warm soapy water between each
patient's use. The patients should be instructed to clean this filter at least once per week,
depending on the environment, during normal operation.
Change the product filter after 10,000 hours of use. It is not necessary to change this filter
between patients even if the previous patient had a communicable disease or infection.
Allow FreeStyle to air dry, and then re-test oxygen concentration before you return the
unit to your inventory.
*Make the bleach solution a 1:100 dilution of 5.25% sodium hypochlorite. Mix one part
household bleach (e.g., Clorox with 99 parts cold tap water). To measure the solution
easily, take 1/4 cup of household bleach, and mix it with a gallon of cold tap water. Allow
the mixture to sit on potentially contaminated surfaces for
10 minutes.
5.0
Service

5.1 Components

The design of AirSep's FreeStyle Portable Oxygen Concentrator allows for easy access
and removal of most components. This allows you to perform scheduled maintenance and
to repair and replace parts with minimal time and effort.
For your safety, be sure to disconnect the FreeStyle Universal AC / DC adaptor from
the power outlet before you service the concentrator.
Note all scheduled maintenance. (Refer to Section 4.0 of this manual.)
5-1
AirSep does not recommend the sterilization of this equipment.
FreeStyle Service Manual MN126-1 E