Freestyle MN126-1 Service Manual page 49

Table of Contents

Advertisement

9. Remove hour meter as per Section 5.12.
10. Remove cannula nozzle tubing by cutting tie-wrap (use small-size cutter).
Carefully pull tubing from nozzle barb. (Figure 5.32)
11. Disconnect fan connector using needle-nose pliers, carefully pulling upward
on fan connector housing at location J6 on main circuit board.
12. Remove six 4-20 self-tapping plastic screws.
(Figure 5.33)
FreeStyle Service Manual MN126-1 E
Figure 5.32
Figure 5.33
5-23

Advertisement

Table of Contents
loading

Related Products for Freestyle MN126-1

Table of Contents