10. Reassemble FreeStyle, and leak test at bed connections.
5.9

Feed/Waste Manifold Replacement

1. Remove PSA system per Section 5.4.
2. Carefully remove tie-wraps that bundle the solenoid valve wires together.
3. Loosen and remove the center bolt from the PSA assembly using a 5/32"
Allen wrench.
4. Holding both sieve beds, gently pull upward, removing the product manifold
and sieve beds. Set product manifold and sieve bed assembly aside, keeping
top of sieve bed connected to product the product manifold. (Figure 5.26)
FreeStyle Service Manual MN126-1 E
Figure 5.25
Figure 5.26
5-17