To load the probe:
1. Unwrap the probe.
2. Remove the lid of the
probe container as
shown in the figure
above. Point the
tonometer upward.
1X
4. Press the measurement
5. The probe moves rapidly
button (13) briefly (1
back and forth.
second) to activate the
probe.
Icare HOME
English
3. Drop the probe into
the probe base (1)
by turning the probe
container upside down.
LOAD
LOA
LO
MEASURE
EAS
EASU
REPEAT
DONE
SERVICE
BATTERY
6. See that the Measure
light (15) flashes. If so,
the probe is loaded
correctly and ready for
measurement.
10
www.icaretonometer.com
Icare HOME tonometer Instruction manual for health care professionals

7. USING THE TONOMETER

7.1 CHOOSING THE MEASUREMENT MODE

The tonometer can operate in two modes:
Series mode, press the measurement button constantly.
Single mode, press the measurement button briefly
(1 second) for each of the six measurements.
Icare HOME
English
Series mode
The series mode is especially
useful in self-tonometry. In
the series mode, keeping
the button pressed down
(see figure left) initiates the
measurement function, and
the tonometer takes six rapid
measurements one after the
other to obtain the final IOP
reading.
Single mode
You can use the single mode to
take individual measurements
one at a time. The single mode
is especially useful for those
patients who tend to blink
heavily. Here you press the
measurement button briefly
(1 second) for each of the six
measurements to obtain the
final IOP reading (see figure
left).