Icare HOME tonometer Instruction manual for health care professionals

12. CLEANING AND DISINFECTION

NOTE!
Do not carry out any other service procedures. Leave all other service and repairs to the
manufacturer or a certified service center.
You must clean the forehead and cheek supports for each new patient. Use a wipe
dampened with a 70% isopropyl alcohol solution. Do not immerse the tonometer in
water or other liquid. The tonometer must not be immersed or cleaned using too
much water.

13. ACCESSORIES

Part number
Product Description
540
Probe base
TA022-001
Probe base collar
560
Wrist strap
TA022-044
Carrying case
Battery cover
7179
TA022-035
Patient guide
571
Battery 3 V, CR123A
TA022-037
Support position tags
575
USB cable
Icare HOME
English
Weight
Dimensions
4g
7 mm x 38 mm
2g
20 mm x 15 mm
3g
270 mm x 10 mm x 10 mm
210g
270 mm x 135 mm x 60 mm
3g
26 mm x 23 mm x 7 mm
33g
210 mm x 90 mm x 2 mm
17g
17 mm x 35 mm
40g
70 mm x 41 mm x 13 mm
23g
1m
26
www.icaretonometer.com
Icare HOME tonometer Instruction manual for health care professionals

14. TECHNICAL AND PERFORMANCE DATA

Type TA022
Dimensions: approximately 11cm x 8cm x 3cm.
Weight: approximately 150g.
Power supply: 2 x CR123 non-rechargeable batteries (make sure you use batteries with
built-in PTC protection, for example Energizer Lithium Photo 123 3V CR123A).
Measurement range: 5-50 mmHg.
Accuracy (95% tolerance interval relatively to manometry): ±1,2mmHg (<20 mmHg) and
±2.2 mmHg (≥20 mmHg).
Repeatability (coefficient of variation): < 8%.
The serial number is located on the inside of the battery compartment cover.
The lot number of the probes is on the side of the probe box and the blister packing.
There are no electrical connections from the tonometer to the patient.
The tonometer has BF-type electric shock protection.
Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30% to 90%
Atmospheric pressure: 800hPa – 1060hPa
Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10% to 95%
Atmospheric pressure: 700hPa – 1060hPa
Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10% to 95%
Atmospheric pressure: 500hPa – 1060hPa
Environmental restrictions for professional use include:
Medivac vehicles or similar where vibration or noise levels are so high
that the user cannot hear error signals.
Environmental restrictions for lay operators (patients):
Environments where noise is so high that the user cannot hear the error
signals.
Mode of operation: continuous