Icare HOME tonometer Instruction manual for health care professionals

6. SETTING UP THE TONOMETER

Setting up your Icare HOME tonometer is easy, with few steps. The following
subchapters describe how you get started.

6.1 INSTALLING OR CHANGING THE BATTERY

NOTE!
Make sure you use batteries with built-in PTC protection, for example Energizer Lithium
Photo 123 3V CR123A.
Update the tonometer's time to your local time. It is done automatically by performing
the steps 1 and 2 under section 8. Reading the measurement data.
Lift the silicon lid that protects the USB port and keeps
the battery compartment cover in place. Open the
battery compartment cover by pressing the silicon lid
2.
slightly and sliding the battery compartment cover as
shown in the figure left.
1.
1. Silicon lid
2. Battery cover
Opening the silicon lid and battery cover
Insert two CR123A lithium batteries in the correct
order: (+) end upwards as shown in the figure left.
Close the cover firmly and press the silicon lid in place
to cover the USB port.
2 x CR123A
Inserting new batteries

6.2 TURNING THE TONOMETER ON

NOTE!
The tonometer turns off automatically after 3 minutes if you do not use it.
Press the power button (20) to turn the tonometer on. The lights (14-19) are
turned on briefly. Following a brief pause, the Load light flashes on the back
panel to remind the user to load the single-use probe into the tonometer prior to
measurement.
Icare HOME
English
8
www.icaretonometer.com
Icare HOME tonometer Instruction manual for health care professionals

6.3 LOADING THE PROBE

WARNING!
To prevent contamination, do not touch the bare probe, do not use a probe if it touches
a non-sterile surface like a table or a floor.
WARNING!
Use only the original and certified probes made by the manufacturer. The probes are for
single-use (single pair of measurement sequences) only. Use probes taken only from the
intact, original packaging. The manufacturer cannot guarantee sterility of the probe once
the seal is compromised. Re-sterilization or re-use of the probe could result in incorrect
measurement values, in the breakdown of the probe, cross-contamination of bacteria or
viruses, and infection of the eye. Re-sterilization or re-use will void all responsibilities and
liabilities of the manufacturer concerning the safety and effectiveness of the tonometer.
The Icare HOME tonometer uses single-use probes
that are packed in a plastic tube and wrapped in blister
packs as shown in the figures left.
Icare HOME
English
9
www.icaretonometer.com