7.5 Use of Forza
Application and acquisition with the Forza device
To use the Forza system properly, please follow these instructions:
PC use of Forza:
• connect the load cell to the Forza device using the Load Cell Input ;
• connect the device to your PC using CUSB03 cable;
• launch the SW OT BioLab and select the device Forza;
• create the appropriate setup and enter the values of sensitivity and full scale of the load
cell;
• delete offset via SW and proceed with the display of the signal
Tablet and Smartphone use of Forza:
• connect the load cell to the Forza device using the Load Cell Input;
• pair the device Forza to a Tablet or to a Smartphone via Bluetooth;
• launch the application called Forza;
• press connect button from the application ;
• set the values of sensitivity and full scale use of the load cell in the App ;
• delete offset from the App and proceed with the acquisition of the signal
Analog mode use of Forza:
• connect the load cell to the Forza device using the Load Cell Input;
• connect the device to an acquisition system using the cable called J-BNCcable;
• adjust the gain of the system using the button Gain Select;
• delete the offset using the button Offset Null
Formula for the force calculation:
Force (Kg) = [V
(V) *F.S.(Kg)] / [Sensibility (mV/V) * 5 (V) * Gain (V/V) ]
out measured