7.1 Controls and indicators

Description of controls, indicators and connectors shown in fig.1:
- Load Cell Input: This input allow the connection of bridge tranducers like load cells.
- Battery/Error LED: This red LED is normally off, it switch on when the battery is empty
and when there is some loose of data during Bluetooth transmission.
- Status LED: This LED flashes blue when Forza is working.
- Recharge LED: This red LED switches on to indicate that battery are under charge. It
switch off when battery are fully recharged.
- USB connector: this connector is used to recharge battery and to transfer data from
Forza to the PC.
- ON/OFF switch: With this switch is possible to turn off and on the Forza system.
When the system is connected via USB to the PC it is automatically on, independently of
the ON/OFF switch position.
- Offset Null button: this button allows you to reset the output offset of Forza in the analog
mode use (for digital mode use offset reset occurs via SW or via App) .
- Gain Select button: this button allows you to change the system gains in the analog
mode use (digital mode use has fixed gain of 100V/V ) .
- Gain led: this led shows the selected gain with different colors: gain 100 = led off, gain
200 = light green, gain 500 = light blue , gain 1000 = red LED )
- OUT: At this connector is available the analog force signal amplified, filtered and with the
possibility to delete offset (everything is after the optical isolation of the device).