Chamber Temperature Setting; Remote Alarm Terminal - Panasonic MDF-594C Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

CHAMBER TEMPERATURE SETTING

Table 1 shows the basic operation method. Perform key operations in the sequence indicated in the
table. The example in the table is based on the assumption that the temperature is -70
Note: The unit is set at the factory that the chamber temperature is -80
Table 1. Basic operation sequence (Example: Chamber temperature -70
Description of operation
1 Turn on the power switch.
Press mode setting key when
2
changing the set temperature
during operation.
Set the second digit to 7 with the
3
numerical value shift key.
4 Press enter key.
The temperature set mode returns to the temperature display area mode automatically when 90
seconds has passed without any key operation.

REMOTE ALARM TERMINAL

The terminal of the remote alarm is located at rear side of the unit. The alarm is outputted from this
terminal.
Contact capacity is 2 A (DC 30 V) .
a) Contact output: normal open connect with N.O. and COM.
b) Contact output: normal close connect with N.C. and COM.
Consult our dealer or our office for installation of the remote alarm.
Key operated
The current chamber temperature
-----
is displayed.
PV/SV
The second digit is flashed.
When pressed the figure of settable
digit changes.
Set temperature is memorized and
ENT
the current chamber temperature is
displayed.
14
o
C.
o
C.
o
C)
Indication after operation