6 Error: causes and rectifications

Indication
Display empty

7 Maintenance and spare parts

• Use a soft, slightly moistened cloth to
clean the OMRON MX3 Plus.
Caution! Do not use petrol, thinners or
similar solvents!
• Use a soft, moistened cloth and soap to
clean the cuff.
Caution! Do not wash the cuff!
Caution! Do not use petrol, thinners or
similar solvents!
• Do not carry out repairs of any kind
yourself. If a defect occurs consult your
8
Error
Cause
Correct
Cuff pressure
reading
low
not
possible
Movement
Very weak
artery pulse
Cuff pressure
too high
Low
Batteries weak/ Fit four new identical 1.5V
voltage
exhausted
Rectification
Wait for at least three minutes to repeat
measurement.
Inflate the cuff 40 mmHg higher than
previous value.
Follow instructions, repeat
measurement.
Wait for at least three minutes to repeat
measurement.
Inflate the cuff 30 mmHg lower than
previous value.
Follow instructions, repeat
measurement.
alkaline batteries type AA
LR6
OMRON retail outlet or distributor as
mentioned on the packaging.
Calibration
• It is recommended to have the OMRON
MX3 Plus inspected every two years to
ensure correct function and accuracy.
Consult your OMRON retail outlet or
OMRON distributor as mentioned on the
packaging.
Replace batteries
• Remove the battery out of the battery