PROCEDURE 11
This Procedure Includes the Following:
Installing/Removing Batteries into/From Battery
Boxes
Connecting Battery Cables
B
When to Charge Batteries
A
Charging Batteries
T
T
Replacing Batteries
E
Installing/Removing Battery Boxes
R
I
E
S
Make sure power to the wheelchair is OFF be-
fore performing this procedure.
The use of rubber gloves and safety glasses is rec-
ommended when working with batteries.
Invacare strongly recommends that battery in-
stallation and battery replacement always be
done by a qualified technician.
After ANY adjustments, repair or service and BE-
FORE use, make sure all attaching hardware is
tightened securely - otherwise injury or damage
may result.
INSTALLING/REMOVING
BATTERIES INTO/FROM BATTERY
BOXES (FIGURE 1)
NOTE: To remove the battery(ies) from the battery box(es),
reverse the following procedure.
NOTE: Have the following tools available:
TOOL
Battery Lifting Strap
1/2-inch (6 pt) Box Wrench
7/16-inch (6pt) Box Wrench
3/8-inch (6pt) Box Wrench
Diagonal Cutters
*NOTE: The Battery Lifting strap supplied is for Group
22 NF Batteries ONLY. Refer to the battery manufac-
turer for the proper lifting strap and/or battery tools for
U1 battery removal/installation.
ALWAYS use a battery lifting strap when lifting a
battery. It is the most convenient method and
assures that the battery acid will not spill. It also
helps to prolong the life of the battery.
DO NOT tip the batteries. Keep the batteries in
an upright position.
WARNING
QTY
COMMENTS
1
See Note
1
Not Supplied
1
Not Supplied
1
Not Supplied
1
Not Supplied
WARNING
NOTE: If there is battery acid in the bottom or on the
sides of the battery box(es) or battery(ies), apply bak-
ing soda to these areas to neutralize the battery acid.
Before reinstalling the NEW or existing battery(ies),
clean the baking soda from the battery box(es) or
battery(ies).
NOTE: When securing battery lifting strap to battery, ob-
serve polarity markings located on the ends of the battery
lifting strap, (+) side to POSITIVE (+) battery terminal/
post and (-) side to NEGATIVE (-) battery terminal/post.
1. If necessary, remove the battery boxes from the
wheelchair. Refer to INSTALLING/REMOVING BAT-
TERY BOXES in this procedure of the manual.
2. Secure the battery lifting strap to battery terminal(s)/
post(s) (FIGURE 1).
CAUTION
Some battery manufacturers mold a carrying strap
and/or hold down flanges directly into the bat-
tery case. Batteries which interfere with the bat-
tery box cannot be used for these applications.
Attempting to "wedge" a battery into a battery
box may damage the box and/or the battery.
3. Place batteries into battery box bottom.
Battery Lifting Strap
Group 22 NF ONLY
Battery Terminal(s)/Post(s)
Group 22 NF
Battery Box
Bottom
FIGURE 1 - INSTALLING/REMOVING BATTERIES INTO/
FROM BATTERY BOXES
38
BATTERIES
Molded Carrying Strap
Or Hold Down Flanges
Battery
U1 Battery Box
Bottom

Advertisement