CEFAR MYO X2 & CEFAR ACTIV X2
Number of channels . . . . . . . . . . . . . . . . . . . . . . . . . .2
Constant current . . . . . . . . . . . . . . . . . . . . . . . . . . . .Up to a resistance of 1000 ohms
1
Stimulation current/channel . . . . . . . . . . . . . . . . . . . .0–100 mA
Waveform. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Symmetrical biphasic pulse,
Number of preset programs
CEFAR MYO X2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
CEFAR ACTIV X2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Stimulation forms . . . . . . . . . . . . . . . . . . . . . . . . . . . .Continuous stimulation:
Max pulse duration . . . . . . . . . . . . . . . . . . . . . . . . . . .350 μs
Max frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80 Hz
Environment for storage, use and shipping . . . . . . . . .Temperature 10 º C–40 ºC
Power source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 x 1.5 V AA non-rechargeable
Current consumption for one
channel, 80 Hz, 30 mA. . . . . . . . . . . . . . . . . . . . . . . .150 mA
I r.m.s. max/channel . . . . . . . . . . . . . . . . . . . . . . . . .11 mA
Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120 x 50 x 30 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ca. 180 g
2526_CEFAR_MYO.ACTIV_X2__ENG_3-114 14

KEY TO ThE STIMULATOR SYMBOLS

Read the operating instructions before to use.
100 % compensated
Patient part type – Body Floating.
Conventional
Dispose of the worn-out stimulator in accordance with local and national
Modulated pulse duration
regulations.
Modulated frequency
Intermittent stimulation:
One ore more of the following markings may appear on your device:
0413
Air humidity 30 %–75 %
Complies with the European Medical Device Directive (93/42/EEC).
Air pressure 700 hPa–1060 hPa
Notified body Intertek ETL Semko (0413).
or 2 x 1.2 V AA rechargeable
batteries
Complies with UL 69050-1, SCA C22.2 No. 69050-1.
Certification mark issued by SGS.
The stimulator has a two-year warranty against defects,
on condition that it is used correctly.