Pioneer MJ-D707 Service Manual page 43

Hide thumbs Also See for MJ-D707:
Table of Contents

Advertisement

QQ
3 7 63 1515 0
6.2.2 Test Mode 2 (MD Mechanism Assy Adjustment, Check)
1
Temperature Check (Please perform this check soon after the power has been switched on.)
Note: When IC101 (RF-AMP) or IC104 (mechanical controller) has been exchanged or when a correction changing VD+3 has been made,
be sure to correctly perform all steps up to step 7.
Step No.
Operation Keys and Operation Method
1
Power ON while holding the ENTER key and the
PLAY key depressed.
2
Press the NAME CLIP key.
3
When
is OK (step 4 and following can be omitted).
When
is 80 (not passed)
4
Measure the ambient temperature T (°C) andcheck
hat the measuring value (
in regard to the microcomputer set value (
Note: When the indication of the measuring value
(
) is not stable, and the ambient temperature
T (°C) is 25 °C or less, set to the lower value.
When the ambient temperature is over 25 °C, set to
the higher value.
5
Change
A-B key (DOWN).
TE
L 13942296513
6
Press the EDIT/NO key.
2
Laser Power Check
Step No.
Operation Keys and Operation Method
1
Power ON while holding the ENTER key and the
PLAY key depressed.
2
Press the NAME key.
3
Measure the laser power with an optical power meter.
4
At this time, check that the measuring value (PLAY
power) meets the standard value of the table on the
right.
Note: When the standard value on the right is applied,
the numerical value immediately after entry into test
mode 2 shall be taken as the standard.
5
Press the NAME key.
6
At this time, check that the measuring value (REC
power) meets the standard value of the table on the
right.
www
.
7
Press the STOP key.
http://www.xiaoyu163.com
is 80 or more at this time, the adjustment
) is within the standard
with the TIME SKIP key (UP) and the
x
ao
u163
y
i
http://www.xiaoyu163.com
2 9
8
FL Display
EJECT
Test mode
TEMP
The microcomputer starts measuring.
End
TEMP
8 0
80 is the default value (unadjusted).
TEMP
µ-com Measuring value
).
Room Temperature
[ T (°C) ]
11.6 to 16.9
17.0 to 22.3
22.4 to 27.6
27.7 to 33.0
33.1 to 38.4
TEMP 7B 7B
Q Q
3
6 7
1 3
1 5
No Disc
Writing to EEPROM is made and
return is made automatically from
test mode to normal mode.
FL Display
EJECT
Test mode
ppw
PLAY power output status
ppw
ppw
µ-com
Measuring value
hex =
(
+02 to +04
–01 to +01
–02 to –04
rpw
REC power output status
rpw
hex =
(
+02 to +04
co
–01 to +01
–02 to –04
.
The LD goes out.
MJ-D707, MJ-17D
9 4
2 8
Status
µ-com Set value
hex=
(
)
+1 to +3
0 to +2
When the
–1 to +1
microcomputer
value is
0 to –2
–1 to –3
0 5
8
2 9
9 4
2 8
Status
Laser Wavelength:
May be omitted.
TEMP
µ-com
Set value
Standard Value
)
(mW)
0.69±0.12
0.67±0.1
0.63±0.1
Standard Value
)
(mW)
m
6.18±0.93
5.79±0.87
5.47±0.82
9 9
Remarks
: Measuring
value
: Set value
79
7A
7B
7C
+
7D
9 9
Remarks
780 nm
43