Replacement Of Battery - Panasonic MDF-C8V1 Series Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

REPLACEMENT OF BATTERY

Location of a nickel-metal-hydride battery
This unit is provided a nickel-metal-hydride battery for the power failure warning device. The battery is
located at the back of the control panel. (Fig. 1)
The high voltage components are enclosed in the side cover. The side cover should be removed
by a qualified engineer or a service personnel only to prevent the electric shock.
Removal of nickel-metal-hydride battery
1. Turn off the power switch, battery switch and disconnect the power supply plug.
2. As shown in the Fig. 2, remove 1 screws fixing the side cover with a screw driver and remove the side
cover.
3. Disconnect the battery connector. (Fig. 3)
4. Remove 4 screws fixing the battery cover. (Fig. 4)
5. Take out the battery.
Handling of battery
Cover the battery terminal with an insulating tape to avoid the short circuit. Then follow the procedure for
recycling or proper disposal.
Battery
Battery connector
Fig. 1
Fig. 3
23
Side cover
Fig. 2
Battery cover
Fig. 4