TEMPERATURE RECORDER (OPTION)

Installation of MTR-85H

1. Remove 4 screws (A, B, C, D) on the side of the
front panel to take off the front panel. Then remove 4
screws on the cover for the recorder mounting space
and take off the cover (Fig. 1). After removing the
cover, replace 4 screws.
2. As shown in Fig. 2, install the temperature recorder
to the front panel by using the fixture enclosed with the
temperature
recorder
(MDF-S3085).
3. Lead the recorder sensor to the chamber through
the access port at the bottom of the chamber.
Note:
The port is covered with a thermal insulation and
rubber cap.
Remove those covers before passing
the recorder sensor.
4. Attach the sensor cover to the right side of the
chamber with 2 screws on the chamber wall.
5. Fix the recorder sensor to the inside of the sensor
cover (MTR-C8) by using the enclosed clips and
screws and seal the access port on the chamber side
with a silicon. (Fig. 3)
and
the
recorder
fixing
Front panel
A
B
Temperature
recorder
Recorder fixing
Fixture
Recorder sensor
Sensor cover
31
C
D
Cover
Fig. 1
Front panel
Fig. 2
Silicon
Fig. 3