ALARM TEMPERATURE SETTING
Table 4 Procedure for setting low temperature alarm
Description of operation
1
Press numerical value shift key for
2
5 seconds.
Press numerical value shift key and
3
scroll the figure to 2.
4
Press set key.
Set the temperature to -05 with the
5
digit key and numerical value shift
key.
6
Press set key.
Key operated
----
The current chamber temperature is
displayed.
The right digit is flashed.
When pressed, the figure of settable
digit increases.
SET
The right digit is flashed.
Pressing the key shifts the digit which can be set.
When pressed, the figure of settable
digit increases.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
16
Indication after operation