Troubleshooting - Abbott FreeStryle Lite Owner's Booklet

Table of Contents

Advertisement

Message
What It Means
Incorrect test procedure.
For example, putting
blood on the test strip
before inserting it into
the meter or applying
blood before
and
appear
There may be a problem
with the test strip
Meter error
There may be a problem
with the test strip
Meter error
19
ART22315-101_rev-B.indd 19-20
What To Do
1. Wait for
and
to appear before applying
blood or control solution.
2. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
3. If the control solution result is out of range or the
error reappears, contact Customer Care.
1. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
2. If the control solution result is out of range or the
error reappears, contact Customer Care.
www.AbbottDiabetesCare.com

Troubleshooting

What It Means
1. The meter
Test strip is inserted
does not
upside down, or bottom
enter test
end in, or not fully
mode after
inserted into the meter
inserting a
test strip.
No battery is installed
Battery is installed
incorrectly
Dead battery
Defective test strip or
meter
Blood or foreign objects
put in the test strip port
What To Do
1. Insert test strip print side up, top end in.
1. Install battery with (+) facing up.
1. Replace battery. Reset date and time, if
necessary.
2. If meter still does not enter test mode,
contact Customer Care.
1. Contact Customer Care.