Miscellaneous - Medisana 88310 Instruction Manual

Tens pain therapy tdp
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 31
Cleaning and
maintenance
Accessories for
MEDISANA
TDP
Technical
data
• Remove the batteries before you clean the unit.
• Never use strong cleaning agents or stiff brushes.
• Clean the unit with a soft cloth which has been lightly moistened with a mild
soap sud. Never dip the unit into water for cleaning and ensure that no water
gets into the unit if applicable. Only use the unit again, when it is completely dry.
• Electrodes that do not stick anymore can be cleaned under cold running
water. After cleaning, leave the electrodes on the carrier paper for 24 hours
to restore the adhesive strength.
Cable set
4 self-adhesive multi-use electrodes,
30 x 50 mm
2 self-adhesive multi-use electrodes,
50 x 90 mm
Name and model
Power supply
Channels
Curve form
Output
Intensity
Frequency
Pulse width
Operating conditions
Storage conditions
Dimensions (L x W x H)
Weight
Article no.
EAN number
Electromagnetic compatibility:
The device complies with the EN 60601-1-2 standard for electromagnetic com-
patibility. Visit MEDISANA's website www.medisana.de for details on this meas-
urement data.
Safety check (in the event of commercial use):
The device has been calibrated by the manufacturer for a period of two years. A
safety check must be conducted at least every two years if used for commercial
purposes. The check is subject to charge and can be performed by a competent
authority or authorised maintenance services in accordance with the "German
Medical Product Operator Ordinance".

Miscellaneous

article number 88315
article number 87080
article number 87100
MEDISANA TENS pain therapy unit TDP
4 x 1.5 volt, micro-batteries (AAA, LR03)
or equivalent rechargeable batteries
2
symmetrical bi-phasic, mono-phasic
constant current
in 60 steps from 0 to 60 mA
2 – 120 Hz, depending on the program
50 – 400 μs, depending on the program
5 °C to +40 °C at 30 to 75 % rel. humidity
-10 °C to +50 °C at 10 to 90 % rel. humidity
approx. 115 x 67 x 35 mm (including belt clip)
approximately 99 g (without batteries)
88310
4015588 883101
We reserve the right to make technical changes
and changes to the design.
GB