Unusual Test Results

If your test result doesn't match the way you feel, please follow these steps:
1. Run a control solution test, Chapter 2 "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3 "Testing Your Blood glucose."
3. If your test results still don't reflect the way you feel, call your doctor immediately.
1. Extremely high humidity may affect the test results. A relative humidity greater than
90% may cause inaccurate results.
2. A red blood cell count (Hematocrit) that is either very high (above 55%) or very low
(below 30%) may not provide accurate results.
3. Some studies have shown that electromagnetic fields may affect results. Do not test
near an operating microwave oven.