Expected Values

The EasyPlus mini R2N Blood Glucose test strips are whole-blood referenced and calibrated for easier
comparison to lab results. The normal fasting blood glucose range for an adult without diabetes is 70~110
mg/dL. Two hours after meals, the blood glucose range for an adult without diabetes is less than 120 mg/dL.
For people with diabetes: please consult your doctor for the blood glucose range appropriate for you.

Unusual Test Results

If your test result doesn't match the way you feel, please follow these steps:
1. Run a control test, Chapter 2, "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3, "Testing Your Blood glucose."
3. If your test results still don't reflect the way you feel, call your doctor immediately.
Note:
1. Extremely high humidity may affect the test results. A relative humidity greater than 90% may
cause inaccurate results.
2. Hematocrit below 20% may cause higher results. Hematocrit above 60% may cause lower
results.
3. Some studies have shown that electromagnetic fields may affect results. Do not test near an
operating microwave oven.